Activity Aware Pulse Rate Algorithm Project

This project has 2 main parts:

  • Part 1 - Develop a Pulse Rate Algorithm on the given training data. Then Test Your Algorithm and see that it has met the success criteria.
  • Part 2 - Apply the Pulse Rate Algorithm on a Clinical Application and compute more clinically meaningful features and discover healthcare trends.

Let's start with some background.


Introduction

A core feature that many users expect from their wearable devices is pulse rate estimation. Continuous pulse rate estimation can be informative for many aspects of a wearer's health. Pulse rate during exercise can be a measure of workout intensity and resting heart rate is sometimes used as an overall measure of cardiovascular fitness. In this project you will create a pulse rate estimation algorithm for a wrist-wearable device. Use the information in the section below to inform the design of your algorithm. Make sure that your algorithm conforms to the given specifications.

Physiological Mechanics of Pulse Rate Estimation

Pulse rate is typically estimated by using the PPG sensor. When the ventricles contract, the capilaries in the wrist fill with blood. The (typically green) light emitted by the PPG sensor is absorbed by red blood cells in these capilaries and the photodetector will see the drop in reflected light. When the blood returns to the heart, fewer red blood cells in the wrist absorb the light and the photodetector sees an increase in reflected light. The period of this oscillating waveform is the pulse rate.

PPG Mechanics

However, the heart beating is not the only phenomenon that modulates the PPG signal. Blood in the wrist is fluid, and arm movement will cause the blood to move correspondingly. During exercise, like walking or running, we see another periodic signal in the PPG due to this arm motion. Our pulse rate estimator has to be careful not to confuse this periodic signal with the pulse rate.

We can use the accelerometer signal of our wearable device to help us keep track of which periodic signal is caused by motion. Because the accelerometer is only sensing arm motion, any periodic signal in the accelerometer is likely not due to the heart beating, and only due to the arm motion. If our pulse rate estimator is picking a frequency that's strong in the accelerometer, it may be making a mistake.

All estimators will have some amount of error. How much error is tolerable depends on the application. If we were using these pulse rate estimates to compute long term trends over months, then we may be more robust to higher error variance. However, if we wanted to give information back to the user about a specific workout or night of sleep, we would require a much lower error.

Algorithm Confidence and Availability

Many machine learning algorithms produce outputs that can be used to estimate their per-result error. For example in logistic regression you can use the predicted class probabilities to quantify trust in the classification. A classification where one class has a very high probability is probably more accurate than one where all classes have similar probabilities. Certainly, this method is not perfect and won't perfectly rank-order estimates based on error. But if accurate enough, it allows consumers of the algorithm more flexibility in how to use it. We call this estimation of the algorithms error the confidence.

In pulse rate estimation, having a confidence value can be useful if a user wants just a handful of high-quality pulse rate estimate per night. They can use the confidence algorithm to select the 20 most confident estimates at night and ignore the rest of the outputs. Confidence estimates can also be used to set the point on the error curve that we want to operate at by sacrificing the number of estimates that are considered valid. There is a trade-off between availability and error. For example if we want to operate at 10% availability, we look at our training dataset to determine the condince threshold for which 10% of the estimates pass. Then if only if an estimate's confidence value is above that threshold do we consider it valid. See the error vs. availability curve below.

Error vs. Availability

This plot is created by computing the mean absolute error at all -- or at least 100 of -- the confidence thresholds in the dataset.

Building a confidence algorithm for pulse rate estimation is a little tricker than logistic regression because intuitively there isn't some transformation of the algorithm output that can make a good confidence score. However, by understanding our algorithm behavior we can come up with some general ideas that might create a good confidence algorithm. For example, if our algorithm is picking a strong frequency component that's not present in the accelerometer we can be relatively confidence in the estimate. Turn this idea into an algorithm by quantifying "strong frequency component".


Part 1: Pulse Rate Algorithm Overview

Algorithm Specifications

You must build an algorithm that:

  • estimates pulse rate from the PPG signal and a 3-axis accelerometer.
  • assumes pulse rate will be restricted between 40BPM (beats per minute) and 240BPM
  • produces an estimation confidence. A higher confidence value means that this estimate should be more accurate than an estimate with a lower confidence value.
  • produces an output at least every 2 seconds.

Success Criteria

Your algorithm performance success criteria is as follows: the mean absolute error at 90% availability must be less than 15 BPM on the test set. Put another way, the best 90% of your estimates--according to your own confidence output-- must have a mean absolute error of less than 15 BPM. The evaluation function is included in the starter code.

Note that the unit test will call AggregateErrorMetric on the output of your RunPulseRateAlgorithm on a test dataset that you do not have access to. The result of this call must be less than 15 BPM for your algorithm's performance to pass. The test set should be easier than the training set so as long as your algorithm is doing reasonably well on the training data set it should pass this test.

This will be validated through the Test Your Algorithm Workspace which includes a unit test.

Some Helpful Tips

  1. Remember to bandpass filter all your signals. Use the 40-240BPM range to create your pass band.
  2. Use plt.specgram to visualize your signals in the frequency domain. You can plot your estimates on top of the spectrogram to see where things are going wrong.
  3. When the dominant accelerometer frequency is the same as the PPG, try picking the next strongest PPG frequency if there is another good candidate.
  4. Sometimes the cadence of the arm swing is the same as the heart beat. So if you can't find another good candidate pulse rate outside of the accelerometer peak, it may be the same as the accelerometer.
  5. One option for a confidence algorithm is to answer the question, "How much energy in the frequency spectrum is concentrated near the pulse rate estimate?" You can answer this by summing frequency spectrum near the pulse rate estimate and dividing it by the sum of the entire spectrum.

Dataset

You will be using the Troika1 dataset to build your algorithm. Find the dataset under datasets/troika/training_data. The README in that folder will tell you how to interpret the data. The starter code contains a function to help load these files.

Getting Started

The starter code includes a few helpful functions.

  • TroikaDataset, AggregateErrorMetric, and Evaluate do not need to be modified.
  • Use TroikaDataset to retreive a list of .mat files containing reference and signal data.
  • Use scipy.io.loadmat to the .mat file into a python object.
  • The bulk of the code will be in the RunPulseRateAlgorithm function. You can and should break the code out into multiple functions.
  • RunPulseRateAlgorithm will take in two filenames and return a tuple of two numpy arrays--per-estimate pulse rate error and confidence values. Remember to write docstrings for all functions that you write (including RunPulseRateAlgorithm)
  • Finally, run the Evaluate function to call your algorithm on the Troika dataset and compute an aggregate error metric. While building the algorithm you may want to inspect the algorithm errors on more detail.

Part 2: Clinical Application Overview

Now that you have built your pulse rate algorithm and tested your algorithm to know it works, we can use it to compute more clinically meaningful features and discover healthcare trends.

Specifically, you will use 24 hours of heart rate data from 1500 samples to try to validate the well known trend that average resting heart rate increases up until middle age and then decreases into old age. We'll also see if resting heart rates are higher for women than men. See the trend illustrated in this image:

heart-rate-age-ref-chart

Follow the steps in the clinical_app_starter.ipynb to reproduce this result!

Dataset (CAST)

The data from this project comes from the Cardiac Arrythmia Suppression Trial (CAST), which was sponsored by the National Heart, Lung, and Blood Institute (NHLBI). CAST collected 24 hours of heart rate data from ECGs from people who have had a myocardial infarction (MI) within the past two years.2 This data has been smoothed and resampled to more closely resemble PPG-derived pulse rate data from a wrist wearable.3


Citations

  1. Troika - Zhilin Zhang, Zhouyue Pi, Benyuan Liu, ‘‘TROIKA: A General Framework for Heart Rate Monitoring Using Wrist-Type Photoplethysmographic Signals During Intensive Physical Exercise,’’IEEE Trans. on Biomedical Engineering, vol. 62, no. 2, pp. 522-531, February 2015. Link
  2. CAST RR Interval Sub-Study Database Citation - Stein PK, Domitrovich PP, Kleiger RE, Schechtman KB, Rottman JN. Clinical and demographic determinants of heart rate variability in patients post myocardial infarction: insights from the Cardiac Arrhythmia Suppression Trial (CAST). Clin Cardiol 23(3):187-94; 2000 (Mar)
  3. Physionet Citation - Goldberger AL, Amaral LAN, Glass L, Hausdorff JM, Ivanov PCh, Mark RG, Mietus JE, Moody GB, Peng C-K, Stanley HE. PhysioBank, PhysioToolkit, and PhysioNet: Components of a New Research Resource for Complex Physiologic Signals (2003). Circulation. 101(23):e215-e220.

About

No description, website, or topics provided.

Resources

Stars

Watchers

Forks

Releases

No releases published

Packages

No packages published

Languages

  • Jupyter Notebook 100.0%